| Variation ID | Pack Size | List Price | Qty | ||
|---|---|---|---|---|---|
| photo_camera | 240 L x 1 Pack
NE9-240S |
£3608.51 |
Clifton NE9S Digital Lab Oven - 240 Litre
Temperature Range : Ambient 5°C - 300°C
The Clifton Range NE9S Digital Laboratory Ovens by Nickel-Electro offer precision temperature control from ambient +5°C to 300°C with timer functions.
These lab ovens can run at set temperature immediately, use run back timer or enable a delay start feature. Can be configured for gravity convection or with variable fan speed providing forced air circulation within the chamber.
The oven chamber is constructed from high quality stainless steel with high efficiency thermal insulation, contained within an outer steel housing with anti-bacterial paint finish.
With energy efficiency in mind, particular attention is paid to robust design, insulation, a tight door closure with silicone seal and hinge design to maintain chamber temperatures with intelligent temperature control.
Features and Benefits
Safety Features
• Switching on the controls undergoes an automatic diagnostic check.
• Features a temperature sensor fail alarm with error reporting on LED screen.
• Over temperature alarm is automatically set 5°C above set point.
• Door open – heating is interrupted, after a short delay a door sounder chimes until door is closed, heating resumes.
• Fitted with Class 1 oven temperature thermostat for overall unit protection.
Controls
Featuring bold illuminated LED display of measured temperatures in real-time, in increments of ±0.1°C with intuitive to use controls, utilising touch sensitive keypads. Indicators for heating, timer and alarm. Exhaust air flap is dial adjustable from 0 to 100% for chamber venting. Fan chamber convection variable from 0% to 100%. Fan convection improves temperature uniformity and heating recovery time. Door, lockable with 2 point door engagement for an even seal. Left hand side of the oven features a 35mm diameter access port, with perforated flap cover, to reduce heat loss when used with external probes.
Notes
Before ordering please check the room or building access where this unit will be installed and electrical power requirements. Temperature fluctuation measurements are based on temperature readings on the shelf in centre of the chamber. Drying ovens up to 112 litres are stackable. Do not place in direct sunlight or near a heat or cold source. Allow a 100mm air gap away from the wall at rear. Ensure there is a minimum 300mm air gap from top of the unit to ceiling.
All Nickel-Electro products are securely packed for safe delivery. Orders are fulfilled from their Weston Super Mare facility, ensuring fast and reliable UK delivery.
| Product Code | NE9-15S | NE9-32S | NE9-56S | NE9-112S | NE9-240S |
| Chamber Capacity | 15 Litres | 32 Litres | 56 Litres | 112 Litres | 240 Litres |
| Temperature Range | Ambient + 5°C - 300°C | ||||
| Temperature Control | PID Digital, LED display showing actual or set temperature, resolution 0.1°C | ||||
| Temperature Alarm | Over temperature alarm is automatically set 5°C above set point | ||||
| Temperature Sensor | pt100 with control diagnostics at power up | ||||
| Temperature Fluctuation | ±0.4°C at 150°C | ||||
| Temperature Variation | ±1.8°C at 150°C | ||||
| Temperature Heat Up 150°C | 25 Minutes | ||||
| Temp. Door Open Recovery 150°C | 5 Minutes | ||||
| Temp. Control & Run Back Timer | Visual indicator and end messaging 00:01 to 99:59 (hh:mm) activates at set temp | ||||
| Temp Control & Delay Start Timer | Heating commences at a set time in the future, within 24 hours | ||||
| Chamber Convection | Fan speed can be set from 0% gravity convection within the chamber variable to 100% for forced air convection | ||||
| Exhaust Air Flap | Adjustable 0 to 100% chamber venting | ||||
| Access Port | 35mm port size, perforated flap cover helps to reduce losses when fitting independent external probes or thermocouples | ||||
| Doors | Single insulated door, lockable with 2 point door engagement for an even seal | ||||
| Door Alarm | Door open, heating interrupted, delay before door sounder chimes until door closed, heating resumes | ||||
| Shelves Fitted & Max. | 1 wire / 3 max |
2 wire / 5 max | 2 wire / 8 max | 3 wire / 10 max | |
| Shelving Locations | Available every 20mm vertically | ||||
| Shelf Maximum Load | 10Kg | 20Kg | |||
| Heater Power (220-240V) | 1050W | 1100W | 2150W | 3200W | |
| Voltage | 230V 50-60Hz |
230V 50/60Hz blue power plug 2P+E |
|||
| Stackable | Yes | No | |||
| Feet/Castors | Levelling feet | ||||
| Dimensions | |||||
| External (mm) | 520w x 390d x 600h | 600w x 460d x 680h | 600w x 605d x 720h | 600w x 660d x 980h | 800w x 720d x 1160h |
| Internal (mm) | 320w x 180d x 240h | 400w x 250d x 320h | 400w x 400d x 360h | 400w x 450d x 620h | 600w x 510d x 800h |