| Variation ID | Pack Size | List Price | Qty | ||
|---|---|---|---|---|---|
| photo_camera | 1 Pack
HI-2640-02 |
£525.00 |
HI-2640 benchtop meter enables fast, accurate measurements utilising the Hanna Instruments digital electrode / probe for HI-764080 dissolved oxygen. Each digital electrode / probe has a unique serial number and the parameter that it measures is automatically identified by the meter.
Once connected, the single sensor or sensors are ready to measure the parameter along with temperature. The HI-2640 measures dissolved oxygen in ppm and % saturation.
The intuitive design simplifies configuration, calibration, measurement.
The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle.
Digital probe input - simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.
Clear, full text readout - The HI-2640 presents clear, detailed text prompts at the bottom section of the screen.
Atmospheric pressure sensor - To make measurements the most accurate, the dissolved oxygen digital probe used with HI-2640 features an atmospheric pressure sensor for automatic pressure compensation.
GLP - The sensor retains data from the most recent calibration, including date, time, and standards utilised.
Capacitive touch keypad - HI-2640 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue.
Easy to read LCD - HI-2640 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’). The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry.
Warranty - Meter has 2 years. 6 months from date of purchase for electrodes and probes
| Dissolved Oxygen - using HI-764080 DO Probe | |
| DO Range | 0.00 to 45.00 PPM (mg/L); 0.0 to 300.0% saturation |
| Resolution | 0.01 ppm (mg/L); 0.1% saturation |
| Accuracy | ±1.5% of reading (± 1 digit) |
| Calibration | one or two point at 0% (HI-7040L solution) and 100% (water-saturated air) |
| Pressure Compensation | Automatic 450.0 to 850.0 mmHg (0.1 mmHg resolution) |
| Salinity Compensation | 0 to 40 g/L (1g/L resolution) |
| Temperature Compensation | 0.0 to 50.0°C) |
| Temperature | |
| Temperature Range | -20.0 to 120.0 °C – temp limits reduced to probe limits |
| Resolution | 0.1°C |
| Accuracy | ±0.5°C |
| Additional Specifications | |
| Probe | HI-764080 digital dissolved oxygen electrode with 3.5mm connector and 1m cable (included) |
| PC Interface | USB-C |
| Storage Interface | USB-C |
| Power supply | USB Type C (5 VDC; 500mA) |
| Environment | 0 to 50˚C; Maximum 95% RH non-condensing |
| Dimensions | 205 x 160 x 77 mm |
| Weight | Approximately 0.85 kg |