APC Pure Logo webp
shopping_cart 0 search menu

Portaclave, Portable Autoclaves, Electric, Manual Controller

Portaclave, Portable Autoclaves, Electric, Manual Controller Group Picture
Excl. VAT Incl. VAT

£3025.00 to £4250.00

To see other pack sizes, select from drop down

Code: Please choose option from drop down above
Revert
  Pack Size List Price Qty  
photo_camera 15L x 1 Pack
ST1528E
£3025.00
photo_camera 22L x 1 Pack
ST2228E
£3650.00
photo_camera 30L x 1 Pack
ST3028E
£4250.00
Revert

Portaclave Portable Autoclaves – Electric, Manual Controller

Portaclave portable autoclaves provide a compact and reliable sterilisation solution for laboratories, educational environments, and field applications. Designed for simple operation with a manual controller, these electric autoclaves offer dependable steam sterilisation without the need for permanent plumbing or complex installation.

Constructed with a durable Grade 316 stainless steel chamber, the Portaclave range combines robust construction with portability. The units operate from a standard single phase power supply and are available in 15 L, 22 L, and 30 L capacities to suit different laboratory sterilisation requirements.

Key Features

Compact Portable Design: Space-saving vertical design allows the unit to be easily moved and used in laboratories, teaching environments, and field locations.

Manual Control Operation: Simple manual operation provides reliable sterilisation without complex digital controls.

Plug & Play Installation: Operates from a standard 230V single phase supply with no plumbing required.

Durable Stainless Steel Chamber: Manufactured from Grade 316 stainless steel for chemical resistance and long-term durability.

Multiple Capacities Available: Available in 15 L, 22 L, and 30 L configurations to accommodate different sterilisation volumes.

Shipped from our Manchester, UK Facility

All autoclaves are securely packed for safe transport and dispatched from our Manchester warehouse to ensure fast and reliable delivery across the UK.

  • sterilising laboratory glassware and small instruments

  • preparing sterile media and culture vessels

  • routine sterilisation tasks in research laboratories

  • educational laboratory sterilisation procedures

  • field or mobile sterilisation where plumbing is unavailable

Capacities and dimensions
Model Name Portaclave 15E Portaclave 22E Portaclave 30E
Model No. ST1528E ST2228E ST3028E
Capacity 15 L 22 L 30 L
Autoclave Weight 19 kg 21 kg 23 kg
Export Case Size 470w x 680h x 470h 470w x 880h x 470h 470w x 880h x 470h
Overall Dimensions 370w x 485h x 390d 370w x 600h x 390d 370w x 730h x 390d
Chamber Dimensions 280 Ø x 245d 280 Ø x 360d 280 Ø x 490d
Water Capacity 2 L 2 L 2 L
1 L Media Bottles 4 4 8
Basket Dimensions 240 Ø x 220d 240 Ø x 220d 240 Ø x 440d
Digital Controller No No No

 

Safe Design Operating Limits
Model Portaclave 15 Portaclave 22 Portaclave 30
Model No. ST1528 ST2228 ST3028
Max working temperature 121°C 121°C 121°C
Min working temperature 7°C 7°C 7°C
Max ambient temperature 28°C 28°C 28°C
Min ambient temperature 11°C 11°C 11°C
Max hydraulic pressure 2.0 Bar 2.0 Bar 2.0 Bar
Max working pressure 1.3 Bar 1.3 Bar 1.3 Bar

 

Site Requirements
Electrical site requirement: 13amp Socket Outlet
Model Portaclave 15 Portaclave 22 Portaclave 30
Model No. ST1528 ST2228 ST3028
Voltage 230V 230V 230V
Phase Requirement 1ph/N 50/60 Hz 1ph/N 50/60 Hz 1ph/N 50/60 Hz
Power Consumption 2kw 2kw 2kw
Fuse Rating 13 amp 13 amp 13amp

 

Chamber Material: Grade 316 stainless steel

Design: Robust space saving portable design

Installation: Plug & play operation with no plumbing needed

Maintenance: Minimal maintenance - Recommended every 12 months

Water Requirement: Use regular mains water only. Do not use distilled or RO water

Sodium Chloride Limit: Must not exceed 500 ppm

Compliance: PED Cat 1 & 2 Module D1, EMC BS EN 61326, PED 2014/68/EU, PD5500:2022, BS2646:2021, PM73 lid locking devices, ISO 9001:2015, CE marked